The High & Low Temperature ICI type,
Cone & Plate

Analogue Viscometer

Specifications

Electrical supply
220-240 Volts or 110-120 Volts 50/60 Hz single phase.
Power
125 Watts
Standard viscosity range
0-2 poises at 10,000 sec-1
0-2.5 poises at 10,000 sec-1
0-5 poises at 10,000 sec-1
0-10 poises at 10,000 sec-1
0-20 poises at 2,500 sec-1
0-40 poises at 2500 sec-1
0-100 poises at 2500 sec-1
Cone Type
To suit range: 0-2, 0-2.5, 0-5, 0-10, 0-20, 0-40, 0-100 poises
Accuracy
+/- 2% of full scale, confirmed using PRA oils traceable to National Standards.
Warm-up from cold
15 minutes
Temperature control of sample
One or three switched temperatures in the range 10 degrees C - 75 degrees C, six temperatures in the range 25 degrees C - 230 degrees C.
Time for sample to obtain control temperature
20 seconds
Ambient Temperature H.T. Instrument
To calibrate the high temperature instrument at 25 degrees C, the ambient temperature should be <20 degrees C.
Ambient Temperature L.T. Instrument
The ambient temperature will affect the instrument lowest temperature. Lowest temperature achieved is 10 degrees C. below ambient.